The fluorometer is a permanently installed device aboard Thalassa. Instrument used : TURNER 10-AU-005-CE The fluorometer is connected onto the same sea water circuit as the thermosalinometer and measures the concentration of chlorphyll present in the sea water. The phytoplancton cells are lit by a beam of ultraviolet light. Their chlorophyll content absorbs the light which is then re-emitted in infrared. The phytoplancton density can be determined in this way. Technical specifications of the instrument : Sensitivity : 30 parts per trillion of extracted chlorophyll
a Automatic range selection : manual or automatic range changing in response to different concentration levels (user selectable). Ranges : 3 ranges, each a factor of 10 more sensitive than the next from 0 to 9999,999 florescence signal units. Blanks : reads and subtracts blanks (user can select manual or automatic blank correction). Digital output : 100% ASCII format through an RS-232 serial cable Analog output : full scale voltage: 0,1, 1, 2 or 5 volts (user selectable). Readout : direct concentration or raw fluorescence.
|
|
|